303 & 603 Deflection Temperature and Vicat Testers
The Models 603 HDTM and 303 HDTM are the latest generation of digitally controlled Automatic Deflection Temperature / Vicat test equipment. The 603 HDTM is capable of testing up to six different specimens simultaneously, the smaller 303 HDTM can test up to three specimens simultaneously, both with an automated testing sequence that proceeds according to user defined control and configuration parameters.
The basic Model 603 HDTM includes a six station bath, two test stations, and a handheld terminal, and up to four more stations can be added. The basic Model 303 HDTM includes a three station bath, two test stations, and a handheld terminal. Each test frame can be configured with optional accessories for either Vicat or deflection temperature testing, including 4” or 100 mm edgewise or 64 mm flatwise deflection temperature. Additionally, the machines are ready to be linked to Tinius Olsen’s HDV software so that the PC can configure the controller, collect the test data, generate a test report, and save the results.
All test stations are pneumatically raised from the bath at the touch of a button on the handheld terminal, allowing easy placement of the test specimens. Once the specimens are loaded into the respective test positions, all stations are lowered collectively into the oil bath and the test can start.
All testing hardware is fully complemented by developed software that turns your hardware into a sophisticated quality analysis system.
Specifications
Model | 603 | 303 | |
Max. Number of Sample Stations | 6 | 3 | |
Temperature Range | 23°C to 300°C | 23°C to 300°C | |
Temperature Ramp | 50°C or 120°C per hour | 50°C or 120°C per hour | |
Temperature Display Resolution | 0.1 °C | 0.1 °C | |
Temperature Sensor | Platinum RTD located adjacent to the load application point at each station | Platinum RTD located adjacent to the load application point at each station | |
Deflection / Penetration Measurement | LVDT | LVDT | |
Deflection / Penetration Display Resolution | 0.001 mm | 0.001 mm | |
Cool-down Rate | Max 20 °C above cooling water temperature in 20 minutes | Max. 20 °C above cooling water temperature in 30 minutes | |
Temperature Safety Limit | Independent dual systems using thermostatic switch in bath and keypad selectable software limiting | Independent dual systems using thermostatic switch in bath and keypad selectable software limiting | |
Dimensions W x D x H | mm | 1067 x 762 x 572 | 813 x 635 x 585 |
in | 42 x 30 x 22.5 | 32 x 25 x 23 | |
Weight (Net) | kg (lb) | 132 (290) | 87 (190) |
Required Utilities | Heat Transfer Medium | 18 liters (4.8 US gallons) min | 11 liters (2.9 US gallons) |
Water | Water supply for cooldown | Water supply for cooldown | |
Clean Air | Dry air filtered to 50 microns at 40psi (2.7bar) min | Dry air filtered to 50 microns at 40psi (2.7bar) min | |
Power | 220 +/- 10%, 50/60 Hz, 1 phase, 4.5kW | 220 +/- 10%, 50/60 Hz, 1 phase, 4.5kW |